XL-3604 Wideband Current Transformer Calibrator
The XL-3604 wideband current transformer calibrator is mainly used for the calibration of wideband current transformers. Calibration items include ratio error, phase error, etc.
Product Description
Product Introduction
XL-3604 Wideband Current Transformer Calibrator Mainly used for the calibration of wideband current transformers. Calibration items include ratio error, phase error, etc.
Main Functions
- Supports electronic transformers with small signal output
- Supports calibration of traditional current transformers 1A, 5A and voltage transformers 57.7V, 100V.
- Supports testing the harmonic transfer characteristics of capacitive voltage transformers according to the requirements of "T/CEC 296-2020 Capacitive Voltage Transformer Harmonic Transfer Characteristic Test Technical Specification", supporting calibration of fundamental wave, harmonics (2~50th), and interharmonics (0.1~49.9th).
- Supports calibration of electronic transformers that comply with IEC61850-9-2/LE standard.
Main Features
- Wideband current sensor error (ratio error, phase error) test at DCHz~2.5kHz frequency;
- Power frequency error (ratio error, phase error) test of current sensor;
- 2~40th harmonic error (ratio error, phase error) test of current sensor;
- 2~40th harmonic error (ratio error, phase error) influence test of current sensor;
- Analysis of harmonic content and waveform distortion rate of the secondary output of current sensor, and testing of ratio error and phase error of each harmonic.
Technical Parameters
1. Accuracy Class | Class 0.05 |
2. Measurement Accuracy | |
Fundamental Wave Measurement Accuracy | |
Current Measurement Accuracy | 50mA~6A: 0.05% RD; |
10mA~50mA: 0.1% RD; | |
Standard Input Range | 0~6A (5A standard terminal) |
0~100 (100A standard terminal) | |
Tested Input Range | 0~6A |
Ratio Error Measurement Accuracy | 30Hz~200Hz: <0.05% |
200Hz~1kHz: <0.05% | |
1kHz~2.5kHz: <0.1% | |
Phase Error Measurement Accuracy | 30Hz~200Hz: <2′ |
200Hz~1kHz: <5′ | |
1kHz~2.5kHz: <10′ | |
DC Measurement Accuracy | 0.10% |
Current Input Terminal Impedance | Not greater than 0.1Ω (AC mode), not greater than 5Ω (DC mode) |
Measurement Frequency Range | DC~2.5kHz |
3. Human-Machine Interface | |
Display | TFT LCD resolution 800*600 |
Operation | Touch screen operation, industrial computer has USB port, can connect USB mouse and keyboard |
Dimensions | 470mm×330mm×200mm |
Total Weight | 10kg |
4. Environmental Conditions | |
Power Supply Voltage | AC 220V±10% 50 Hz |
Ambient Air Temperature | (-10℃ ~ 55℃) |
Maximum Relative Humidity | <85% |