XL-809T Distribution Network AC Sensor Calibration Device

The XL-809T1 distribution network AC sensor calibration device can perform fully automatic detection of ratio error and phase error for sensors of grade 0.2S and below, automatically generate test reports, and can test voltage or current sensors. It has high work efficiency and is especially suitable for sensor manufacturing enterprises and inspection agencies conducting batch inspections.

Product Description

Product Introduction

XL-809T1 Distribution Network AC Sensor Calibration Device This device organically combines the calibrator with standard current transformers, current boosters, standard voltage transformers, voltage boosters, load boxes, and other equipment (components) into a desktop distribution network AC sensor calibration device. It can perform fully automatic detection of ratio error and phase error for sensors of 0.2S grade and below, automatically generate test reports, and can test voltage or current sensors. It has high work efficiency and is especially suitable for sensor manufacturers and inspection agencies conducting batch inspections.

 

Main Functions

  • Realizes the calibration of ratio error and phase error for distribution network current sensors (ECT), voltage sensors (EVT), and current-voltage sensors (ECVT).
  • The distribution network AC sensor device integrates traditional equipment (components) into one, equipped with current boosters, standard current transformers, self-boosting standard voltage transformers, and AC variable frequency power supply. It has a high degree of automation, and the measurement methods comply with current transformer calibration regulations. All equipment (components) within the device can be individually sent for inspection according to existing calibration regulations. The device is user-friendly for operators to perform testing in laboratories and other locations, providing great convenience.
  • Displays primary voltage, current, frequency, and other related information with full Chinese operation prompts, making operation convenient. It can store multiple sets of measurement data for long-term retention, facilitating comprehensive evaluation of calibrated parameters.
  • Can display the output waveform of the calibrated sensor in real time, along with related parameters, and automatically save calibration data.
  • Allows editing of test plans for customized calibration of test points, conveniently records calibration point error records, and by clicking report generation, specified error data can be saved as Excel files for easy archiving or error analysis.
  • This product has strong adaptability and scalability, supporting subsequent upgrades.

 

Technical Specifications

1.    Calibrator Technical Specifications
Input Range of Tested Small Signal 0 ~ 20V
Accuracy Grade 0.05 Grade
Ratio Error < 0.05% , Phase Error < 2′ (Tested Small Signal ≥50mV )
Ratio Error < 0.1% , Phase Error < 4′ (Tested Small Signal ≥22.5mV )
Standard Voltage Range 20V100V
Standard Current Range 0.05A0.1A1A5A
Input Range of Tested Small Signal 62.5mV250mV1V8V20V
Standard Voltage Transformer Ratio 10kV:100V (10kV/ 3):(100V/ 3)
2.    Standard Current Transformer Specifications
Secondary Current 1A
Frequency 50Hz
Standard Current Transformer Ratio 20A:1A 500A:1A 1000A:1A
Accuracy Grade 0.02S Grade
3.    Current Booster Specifications
Primary Current 5 ~ 1000A
Frequency 50Hz
Power Supply Voltage 0 ~ 250V
Capacity 2000VA
4.    Self-boosting Standard Voltage Transformer Specifications
Primary Voltage 10 kV / 10kV/ 3
Secondary Voltage 100V / 100V/ 3
Frequency 50Hz
Accuracy Grade 0.02 Grade ( 100V ) / 0.02 Grade ( 100/ 3 )
Rated Load 0.25VA ( 100V ) / 0.07VA ( 100/ 3 )
Power Factor 1
Capacity 200VA
5.    Built-in AC Variable Frequency Power Supply Specifications
Capacity 3000VA
Voltage Range 0 ~ 150V AC / 0 ~ 300V AC AUTO
Frequency 45 ~ 100Hz ( 0.01Step )
Load Voltage Regulation Rate 1%
Voltage Adjustment Rate 0.10%
Load Regulation Rate 1%
Frequency Stability 0.01%
Waveform Distortion T.H.D 2% Pure Resistive Load
Display Translation Voltage Vrms , Current Arms , Frequency Fre , Power Wattage , Power Factor PF
Measurement Accuracy Voltage 0.5%FS+5dgt
Current 0.5%FS+5dgt
Frequency 0.01%FS+5dgt
Power 0.5%FS+5dgt
6.    Other Indicators
Operating Power Supply Voltage Range AC 220V±5% 50Hz
Power Consumption ≈3000VA
Preheating Time < 15 Minutes
Operating Temperature -10℃ ~ 45℃
Relative Humidity ≤85% Non-corrosive Gas

Product Document

File Name

Update Time

Related Products

XL-809CT3 Distribution Network Current Sensor Calibration System


XL-809CT3 Distribution Network Current Sensor Calibration System: Integrates the calibrator, standard current transformer, current booster, load box, and other equipment (components) into an organic whole, forming a desktop distribution network current sensor calibration device.

XL-809PT3 Distribution Network Voltage Sensor Calibration System


The XL-809PT3 distribution network voltage sensor calibration system can perform fully automatic testing of voltage sensors with a grade of 0.2S and below, automatically generate test reports, and test 3 voltage sensors at a time. It has high work efficiency and is especially suitable for voltage sensor manufacturers and inspection agencies conducting batch testing.

XL-809A Current Transformer Automatic Testing Device


The XL-809A current transformer automatic testing device is a verification device developed by our company that integrates a primary large current source output and current transformer calibration. It is mainly used for automatic testing devices of electronic current transformers in digital substations.