XL-809A Current Transformer Automatic Testing Device

The XL-809A current transformer automatic testing device is a verification device developed by our company that integrates a primary large current source output and current transformer calibration. It is mainly used for automatic testing devices of electronic current transformers in digital substations.

Product Description

Product Introduction

XL-809A Current Transformer Automatic Testing Device This is a verification device developed by our company that integrates a primary large current source output and current transformer calibration. It is mainly used for automatic testing devices of electronic current transformers in digital substations. The built-in current source is a new generation high-precision standard power source composed of a 1.2G MAC-based DSP, large-scale FPGA, high-speed high-precision DA, and high-fidelity power amplifier. The built-in transformer calibrator also adopts a dual-CPU architecture with powerful functions, greatly expanding bandwidth and improving measurement accuracy through a 24-bit AD chip and 512 times oversampling technology. The XL-809A itself has an accuracy level of 0.05 and can calibrate current transformers below the 0.2S level.

 

Features

  • Can output a pure sinusoidal standard current source signal with a distortion rate of 0.03% (typical value).
  • Frequency output adjustable from 40Hz to 65Hz with a resolution of 0.002Hz and accuracy of 0.002Hz.
  • Strong load capacity, capable of fully loading capacitive, inductive, resistive, or composite loads, with a load regulation rate better than 0.01% RG.
  • Able to simulate the standard current output of traditional primary transformer measurement; the standard current source output range is 0 to 120A, with a continuous 100A current output time within five minutes, providing a standard current source signal for traditional transformer verification.
  • Calibrator test range is 0 to 5A, meaning the rated secondary current of the tested current transformer can be set arbitrarily within 0.005A to 5A and allows detection up to 120% of the rated value.
  • System comprehensive measurement accuracy better than 0.05%.
  • The calibrator can be configured with a standard transformer and uses a comparison measurement method for testing. Under this test condition, fully automatic testing can be completed with one key using the built-in standard current source, or manual testing can be performed with an external current amplifier.
  • The calibrator can be used without a standard transformer, using the built-in standard current source and direct measurement method for testing. Under this condition, fully automatic testing can be completed with one key.
  • The detection points of the calibrator can be set independently, and the qualification standards can also be set independently. During testing, it can automatically capture test points and determine test results.
  • Realizes calibration of angle difference and ratio difference of traditional transformers, can record test results, conveniently record verification point errors, and generate reports with one click to save recorded error data as Word documents for easy archiving or error analysis. Can output to a computer with options to store or store and print; print format is programmable.
  • Local functions: equipped with an 800*600 large touchscreen LCD and industrial control computer operation, facilitating local operation, software upgrades, and data transfer.
  • Integrated design with compact size and lightweight, convenient for on-site testing work, supporting both touchscreen and keyboard input, providing good usability.

 

Technical Specifications
 

Standard AC Current Source Output Specifications
AC Current Output
Adjustment Fineness 0.01% RG
Accuracy Better than ±0.05% RG
Stability Better than ±0.01% RG/1min
Distortion Better than 0.1% (non-capacitive load)
Output Power 80VA
Full Load Regulation Rate Less than ±0.01% RG
Full Load Regulation Time Less than 1ms
Output Range 0 to 120A
Gear Settings 0A to 0.2A, 0.2A to 1A, 1A to 6A, 6A to 20A, 20A to 100A
Temperature Drift ±8 PPM/°C
Long-term Stability ±60 PPM/year
Frequency Output
Adjustment Range 40Hz to 65Hz
Resolution 0.002Hz
Accuracy ±0.002Hz
Temperature Drift ±0.5 PPM/°C
Long-term Stability ±4 PPM/year
Harmonic Output
Harmonic Order Adjustable from 2nd to 50th harmonic
Maximum Harmonic Content 40%
Transformer Measurement Input
Standard Transformer Input Current
Measurement Accuracy 0.05% RD
Accurate Measurement Range 0 to 6A
Current Measurement Gears 0.2A, 1A, 5A
Tested Transformer Input Current (Large Signal Part)
Measurement Accuracy 0.05% RD
Accurate Measurement Range 10mA to 6A
Current Measurement Gears 0 to 50mA, 50mA to 100mA, 100mA to 1A, 1A to 5A
Tested Transformer Input Current (Small Signal Part)
Measurement Accuracy 0.05% RD
Accurate Measurement Range 0~12mA
Operating Power Supply
Operating Power Supply 220V (±5%) AC @50Hz
Environmental Conditions
Operating Temperature 0℃~40℃
Relative Humidity ≤85%
Storage Conditions -30℃~60℃

Product Document

File Name

Update Time

Related Products

XL-809T Distribution Network AC Sensor Calibration Device


The XL-809T1 distribution network AC sensor calibration device can perform fully automatic detection of ratio error and phase error for sensors of grade 0.2S and below, automatically generate test reports, and can test voltage or current sensors. It has high work efficiency and is especially suitable for sensor manufacturing enterprises and inspection agencies conducting batch inspections.

XL-809CT3 Distribution Network Current Sensor Calibration System


XL-809CT3 Distribution Network Current Sensor Calibration System: Integrates the calibrator, standard current transformer, current booster, load box, and other equipment (components) into an organic whole, forming a desktop distribution network current sensor calibration device.

XL-809PT3 Distribution Network Voltage Sensor Calibration System


The XL-809PT3 distribution network voltage sensor calibration system can perform fully automatic testing of voltage sensors with a grade of 0.2S and below, automatically generate test reports, and test 3 voltage sensors at a time. It has high work efficiency and is especially suitable for voltage sensor manufacturers and inspection agencies conducting batch testing.