XL-847C DC Current Transformer Calibration Instrument Overall Verification Device
XL-847C DC Transformer Tester Overall Calibration Device, used for error calibration of DC transformer testers with analog input based on FT3 input.
Product Description
Product Introduction
XL-847C DC Transformer Tester Overall Calibration Device , Used for based on FT3 Error calibration of DC transformer tester with analog input of DC transformer tester 。
Features
- Calibrate the ratio error of DC transformer tester.
- Calibrate the ratio error of small signal DC transformer tester.
- Calibrate the ratio error of DC transformer tester based on small signal input standard.
- Using standard DC analog source and standard DC digital source output methods, when calibrating digital DC transformer testers, digital micro-differences can be superimposed on the standard DC digital source. Used to verify the DC transformer tester's error detection capability.
- Internally uses a highly stable standard DC analog source, capable of outputting high-accuracy DC voltage, DC current, and small DC voltage signals.
- When superimposing digital micro-differences, based on the FT3 protocol characteristics, a reference value fine-tuning method is used to make the actual digital and analog micro-differences consistent with the set micro-difference value. This solves the numerical quantization problem caused by relative value representation in protocols like FT3.
Technical Specifications
| 1) Accuracy Class | |
| Accuracy Class | Class 0.01 |
| 2) Communication Protocol | |
| Digital Source Output Protocol | IEC60044-8(FT3), IEC61850-9-2LE |
| 3) Reference DC Voltage | |
| Output Range | 7.5mV to 120V |
| Linearity | 20ppm |
| Stability | 0.005%/min |
| Ripple Factor | <0.5% |
| Load Capacity | 20mA (MAX) |
| 4) Reference DC Current | |
| Output Range | 0 to 10A |
| Linearity | 20ppm |
| Stability | 0.005%/min |
| Ripple Factor | <0.5% |
| Load Capacity | Port Voltage 5V (MAX) |
| 5) Reference DC Small Signal | |
| Output Range | 0 to 15V |
| Linearity | 20ppm |
| Stability | 0.005%/min |
| Ripple Factor | <0.5% |
| Load Capacity | 20mA (MAX) |
| 6) Digital Interface | |
| SC Fiber Interface | 1 (Multimode fiber wavelength 1310nm) |
| ST Fiber Interface | 1 (Multimode fiber wavelength 1310nm) |
| RJ45 Interface | 2 |
| FT3 (TX) | 1 |
| FT3 (RX) | 1 (Spare) |
| Output Frame Jitter | <1us |
| 7) Synchronization (Spare Interface) | |
| Optical Synchronization Output Port | 1 (Configurable optical PPS and optical IRIG-B code) |
| Electrical Synchronization Output Port | 1 |
| Optical Synchronization Input Port | 1 (Configurable optical PPS and optical IRIG-B code) |
| Electrical Synchronization Input Port | 1 |
| 8) Human-Machine Interface | |
| Display | TFT LCD Resolution 800*600 |
| Operation | Touch operation, can also connect USB mouse and keyboard for control |
| 9) Others | |
| Total Weight | 16.5kg |
| Dimensions | 570mm×480mm×190mm |
| Power Supply Voltage | 220V±10%, 50Hz |
| External Air Temperature | -10℃ to 55℃ |
| Maximum Relative Humidity | < 85% |