XL-809RE Three-Phase AC Sensor Calibrator

The XL-809RE three-phase AC sensor calibrator is suitable for laboratory and field accuracy calibration of electronic transformers with small signal outputs that comply with IEC60044-7/8 standards. It has an accuracy level of 0.05 and can calibrate electronic transformers with small signal output types of 0.2S level and below.

Product Description

Product Introduction

XL-809RE Three-Phase AC Sensor Calibrator The core adopts a dual-CPU architecture with powerful functions, greatly expanding bandwidth with a 24Bit AD chip and 512 times oversampling technology, improving measurement accuracy. At the same time, multi-gear automatic switching and FIR filters enhance dynamic range and signal-to-noise ratio. Developed with LabView management software, the interface is beautiful and easy to use. The power supply and sampling input adopt extensive electromagnetic compatibility measures, capable of adapting to complex field test environments. The portable design scheme facilitates laboratory and field testing. Suitable for laboratory and field precision calibration of electronic transformers with small signal output conforming to IEC60044-7/8 standards. The XL-809R distribution network AC sensor calibrator has an accuracy of 0.05 level and can calibrate electronic transformers with small signal output types of 0.2S level and below.

 

Main Features

  • Calibration of ratio error and phase error for electronic transformers with small signal output and traditional electromagnetic transformers;
  • Displays current primary voltage, current, frequency, and other related information for comprehensive evaluation of the calibrated parameters;
  • System reference protocols include IEC60044-7/8, JJG313-2010, JJG314-2010, GB20840.7/8-2007;
  • Real-time display of the output waveform of the calibrated sensor and related parameters, with automatic saving of calibration data;
  • Convenient recording of verification point errors; click to generate reports to save recorded error data for archiving or error analysis;
  • Portable design, small size and light weight, convenient for testing work, supports both touchscreen and keyboard input, providing good usability;
  • Strong adaptability and scalability, supporting future upgrades;
  • The input range for the small signal under test is 0~8V, and the zero-sequence small signal voltage input range is 0~20V.

 

Technical Specifications

 

1. Scope of Application
Usage Range Suitable for calibration of single-phase and three-phase electronic and electromagnetic transformers of 0.2S level and below.
2. Standard Voltage Input Channel
Voltage Measurement Ranges 100V, 200V
Voltage Measurement Accuracy (20~240V) 0.05%RD
Ratio Error 0.05%RD
Phase Error 2′ (> 20V)
Input Impedance 260kΩ (100V), 520kΩ (200V)
3. Standard Current Input Channel
Current Measurement Ranges 200mA, 1A, 5A, 10A
Accurate Measurement Range 50mA ~ 10A 0.05%RD
Ratio Error 0.05%RD
Phase Error 2′ (> 10mA)
Input Impedance Less than 0.3Ω @ >200mA, less than 0.2Ω @ >5A
4. Tested Electronic Input Channel (Electronic Voltage Transformer)
Small Signal Voltage Input Range Phase Sequence: 0~8.0V (100mV, 500mV, 2V, 8V)
Zero Sequence: 0~20V (100mV, 500mV, 2V, 20V)
Ratio Error 0.05%RD
Phase Error 2′ (> 22.5mV)
Input Impedance 500MΩ @ 0.6nF
5. Tested Electronic Input Channel (Electronic Current Transformer)
Small Signal Voltage Input Range Phase Sequence: 0~8.0V (50mV, 200mV, 1.25V, 8V)
Zero Sequence: 0~8.0V (50mV, 200mV, 1.25V, 8V)
Ratio Error 0.05%RD
Phase Error 2′ (> 22.5mV)
Input Impedance 500MΩ @ 0.6nF
6. Tested Electromagnetic Input Channel (Electromagnetic Transformer)
Voltage Transformer Input Range 0~240V
Ratio Error 0.05%RD
Phase Error 2′ (> 20V)
Current Transformer Input Range 0~10A
Accurate Measurement Range 50mA~10A 0.05%RD
Ratio Error 0.05%RD
Phase Error 2′ (> 10mA)
7. Frequency Measurement
Frequency Measurement Range 45Hz~55Hz
Frequency Measurement Resolution 0.001Hz
Frequency Measurement Accuracy 0.002Hz
8. Other Specifications
Operating Power Supply Voltage Range AC 220V±5% 50Hz
Power Consumption About 30VA
Warm-up Time < 15 minutes
Operating Temperature 0℃~50℃
Relative Humidity ≤85% No corrosive gases
Dimensions (L×W×H) 365mm×482mm×177mm
Overall Weight 11kg

Product Document

File Name

Update Time

Related Products

XL-809RED Three-Phase Digital Comparative Transformer Calibrator


The XL-809RED three-phase digital comparison transformer tester features a dual-CPU architecture core, offering powerful functionality. It utilizes a 24-bit AD chip and 512 times oversampling technology to greatly expand bandwidth and improve measurement accuracy.

XL-809E Digital Comparison Transformer Calibrator


The XL-809E digital comparative transformer tester is a transformer tester developed for the field of primary and secondary integrated transformer testing in distribution networks. Both the standard and the device under test support multiple types of input signals. It is not limited to testing secondary low-signal transformers but can also test traditional electromagnetic transformers. The testing items include ratio error, phase error, delay, polarity, and more.

XL-809 Electronic Transformer Calibrator


The XL-809 electronic transformer tester (small signal input) is suitable for laboratory and on-site precision calibration of electronic transformers with small signal output that comply with IEC60044-7/8 standards. It has an accuracy level of 0.05 and can perform calibration of electronic transformers with small signal output types of 0.2S level and below.

XL-807H Power Frequency Harmonic Current Transformer Calibrator


The XL-807H power frequency harmonic transformer calibrator supports angle difference and ratio difference calibration for electronic transformers (small signal output) and electronic transformers (IEC61850-9-2/LE output).

XL-809RD Transformer Tester


The XL-809RD transformer tester is suitable for laboratory and on-site precision calibration of electronic sensors with small signal output conforming to IEC60044-7/8 standards and electronic sensors with digital (FT3) output. It has an accuracy level of 0.05 and can perform error verification for electronic sensors of 0.2S level and below.

XL-809RM Three-Phase AC Sensor Calibrator


The XL-809RM three-phase AC sensor calibrator is suitable for laboratory and field precision calibration of electronic transformers with small signal outputs. It has an accuracy level of 0.05 and can calibrate electronic transformers with small signal outputs of 0.2S level and below.

XL-8093I Network Configuration AC Sensor Calibrator


The XL-8093I network distribution AC sensor calibrator (current) has a self-accuracy of 0.05 level and can calibrate current sensors with small signal output types of 0.2 level and below. It features a portable design, convenient for laboratory and field testing.

XL-8093U Network Configuration AC Sensor Calibrator


The XL-8093 distribution network AC sensor calibrator has an inherent accuracy of 0.05 level and can calibrate voltage sensors with small signal output types of 0.2 level and below.

XL-809R Network Configuration AC Sensor Calibrator


The XL-809R network configuration AC sensor calibration instrument is suitable for laboratory and on-site accuracy calibration of electronic transformers with small signal output that comply with IEC60044-7/8 standards. It can perform calibration of electronic transformers with small signal output types of 0.2S grade and below.

XL-3270 Programmable Wideband Current Source


The XL-3270 Programmable Wideband AC Source (hereinafter referred to as XL-3270) can output single-phase AC and DC current. The current output can be sine wave, superimposed DC, DC, half-wave, and various harmonic superpositions. The frequency output range is DC-1.2kHz, suitable for occasions requiring complex power waveform signal output.

Product Inquiry

Our staff will contact you within 24 hours (working days)