XL-809RM Three-Phase AC Sensor Calibrator

The XL-809RM three-phase AC sensor calibrator is suitable for laboratory and field precision calibration of electronic transformers with small signal outputs. It has an accuracy level of 0.05 and can calibrate electronic transformers with small signal outputs of 0.2S level and below.

Product Description

Product Introduction

The XL-809RM three-phase AC sensor calibrator is suitable for laboratory and field precision calibration of electronic transformers with small signal outputs. It has an accuracy level of 0.05 and can calibrate electronic transformers with small signal outputs of 0.2S level and below.

 

  Main Functions

  • Transformer Calibration: Capable of simultaneously calibrating ratio error, phase error, voltage, and current of three-phase electronic transformers.
  • Both voltage transformer calibration and current transformer calibration have 5 input terminals for signals under test, allowing simultaneous calibration of 5 signals. In addition to the three-phase phase sequence test inputs, there are two customizable calibration channels for dual-channel or zero-sequence channel calibration.
  • Energy Calibration: Can perform energy error calibration on pulse inputs using standard voltage and current inputs as references. Suitable for terminals with pulse outputs for calibration.
  • During energy calibration, it can display the amplitude, phase, frequency, active power, reactive power, and power factor of the standard input.
  • Fault Recording Function: Capable of fault recording for three-phase standard input signals, supporting various trigger condition configurations.
  • Supports secondary development protocols and can be integrated with other calibration platforms as a three-phase AC transformer calibrator for secondary development applications.

 

Technical Specifications

 

1. Standard Voltage (Three-phase)
Voltage Measurement Ranges 20V, 100V
Voltage Measurement Accuracy (20–120V) 0.05%RD
Ratio Error 0.05%RD
Phase Error 2′ (> 20V)
Input Impedance 130KΩ
2. Standard Current (Three-phase)
Current Measurement Ranges 200mA, 1A, 5A, 10A
Accurate Measurement Range 50mA ~ 10A 0.05%RD
Ratio Error 0.05%RD
Phase Error 2′ (> 10mA)
Input Impedance 0.7Ω @ ≤100mA, 0.1Ω @ ≤6A
3. Small Signal Voltage Under Test (5 channels)
Input Range 0–8.0V (50mV, 250mV, 1V, 6.25V)
A, B, C Three-phase sequence under test
X, Y Two configurable custom test channels
Ratio Error 0.05%RD
Phase Error 2′ (> 22.5mV)
Input Impedance 200MΩ @ 0.6nF
4. Small Signal Current Under Test (5 channels)
Input Range 0–8.0V (50mV, 250mV, 1V, 6.25V)
A, B, C Three-phase sequence under test
X, Y Two configurable custom test channels
Ratio Error 0.05%RD
Phase Error 2′ (> 22.5mV)
Input Impedance 200MΩ @ 0.6nF
5. Power Measurement and Energy Error
Accuracy 0.05% PF=1 (50mA–10A)
0.1% PF=0.5L, 0.5C (50mA–10A)
Reactive Power Accuracy 0.2% PF≥0.5
6. Energy Pulse Input/Output
Pulse Constant Under Test Decimal numbers can be input; calibration cycle setting range: 1–100, maximum pulse frequency received is 100Hz (units: imp/kWh or imp/kvarh)
Device Pulse Constant Settable range is 3600–2000000 imp/kWh or imp/kvarh; output is optically isolated with load capacity not exceeding 5mA (units: imp/kWh or imp/kvarh)
7. Frequency Measurement
Frequency Measurement Range 45Hz–65Hz
Frequency Measurement Resolution 0.001Hz
Frequency Measurement Accuracy 0.002Hz
8. Other Specifications
Operating Power Supply Voltage Range AC 220V±10% 50Hz±0.5Hz
Power Consumption Approximately 30VA
Warm-up Time <15 minutes
Operating Temperature -10℃ to 55℃
Relative Humidity ≤85% Non-corrosive gases
Dimensions Width 483mm * Height 177mm * Depth 365mm
Total Weight Approximately 11kg

Product Document

File Name

Update Time

Related Products

XL-809RED Three-Phase Digital Comparative Transformer Calibrator


The XL-809RED three-phase digital comparison transformer tester features a dual-CPU architecture core, offering powerful functionality. It utilizes a 24-bit AD chip and 512 times oversampling technology to greatly expand bandwidth and improve measurement accuracy.

XL-809E Digital Comparison Transformer Calibrator


The XL-809E digital comparative transformer tester is a transformer tester developed for the field of primary and secondary integrated transformer testing in distribution networks. Both the standard and the device under test support multiple types of input signals. It is not limited to testing secondary low-signal transformers but can also test traditional electromagnetic transformers. The testing items include ratio error, phase error, delay, polarity, and more.

XL-809 Electronic Transformer Calibrator


The XL-809 electronic transformer tester (small signal input) is suitable for laboratory and on-site precision calibration of electronic transformers with small signal output that comply with IEC60044-7/8 standards. It has an accuracy level of 0.05 and can perform calibration of electronic transformers with small signal output types of 0.2S level and below.

XL-807H Power Frequency Harmonic Current Transformer Calibrator


The XL-807H power frequency harmonic transformer calibrator supports angle difference and ratio difference calibration for electronic transformers (small signal output) and electronic transformers (IEC61850-9-2/LE output).

XL-809RD Transformer Tester


The XL-809RD transformer tester is suitable for laboratory and on-site precision calibration of electronic sensors with small signal output conforming to IEC60044-7/8 standards and electronic sensors with digital (FT3) output. It has an accuracy level of 0.05 and can perform error verification for electronic sensors of 0.2S level and below.

XL-809RE Three-Phase AC Sensor Calibrator


The XL-809RE three-phase AC sensor calibrator is suitable for laboratory and field accuracy calibration of electronic transformers with small signal outputs that comply with IEC60044-7/8 standards. It has an accuracy level of 0.05 and can calibrate electronic transformers with small signal output types of 0.2S level and below.

XL-8093I Network Configuration AC Sensor Calibrator


The XL-8093I network distribution AC sensor calibrator (current) has a self-accuracy of 0.05 level and can calibrate current sensors with small signal output types of 0.2 level and below. It features a portable design, convenient for laboratory and field testing.

XL-8093U Network Configuration AC Sensor Calibrator


The XL-8093 distribution network AC sensor calibrator has an inherent accuracy of 0.05 level and can calibrate voltage sensors with small signal output types of 0.2 level and below.

XL-809R Network Configuration AC Sensor Calibrator


The XL-809R network configuration AC sensor calibration instrument is suitable for laboratory and on-site accuracy calibration of electronic transformers with small signal output that comply with IEC60044-7/8 standards. It can perform calibration of electronic transformers with small signal output types of 0.2S grade and below.

XL-3270 Programmable Wideband Current Source


The XL-3270 Programmable Wideband AC Source (hereinafter referred to as XL-3270) can output single-phase AC and DC current. The current output can be sine wave, superimposed DC, DC, half-wave, and various harmonic superpositions. The frequency output range is DC-1.2kHz, suitable for occasions requiring complex power waveform signal output.