XL-9100T18 DC Meter Calibration Device

The XL-9100T18 DC energy meter calibration device is equipped with 18 independent voltage sources and 1 current source. It can output DC voltage from 0 to 1500V (MAX: 1725V) and DC current from 0 to 100A, enabling incremental output calibration for direct DC energy meters. The voltage accuracy is 0.02 level, current accuracy is 0.03 level, and energy accuracy is 0.05 level. It features a dedicated meter rack with 18 slots, allowing simultaneous calibration of 18 direct energy meters (0~100A).

Product Description

Product Introduction

 

XL-9100T18 DC Energy Meter Calibration Device Built-in 18 independent voltage sources and 1 current source, capable of outputting 0~1500V (MAX: 1725V) DC voltage and 0~100A DC current, enabling incremental output for the calibration of direct DC energy meters. Voltage accuracy is 0.02 level, current accuracy is 0.03 level, and energy accuracy is 0.05 level; designed with a dedicated meter rack with 18 slots, allowing simultaneous calibration of 18 direct energy meters (0~100A).
 

 

Main Features

 

  • The enclosure is made of special aluminum alloy material, elegant and sturdy in structure, lightweight, and easy to transport.
  • Equipped with a dedicated meter rack for convenient meter calibration.
  • High-quality linear power amplifier with low ripple output and good stability.
  • Multiple protection systems with audible and visual alarm functions for higher reliability.
  • The operation software includes shortcut keys for common calibration points, allowing quick setting of frequently used test points during manual meter calibration, making manual calibration more convenient.
  • Equipped with RS485 communication port, capable of communicating with the tested DC meter to read electrical parameters.
  • Fully automatic testing of meter start, creep, basic error, daily timing error, constant test, etc., according to regulations.
  • Supports ripple influence test; voltage and current can superimpose 10~500Hz ripple, with ripple factor and phase adjustable as needed.
  • Dedicated pulse sampling interface with high-voltage isolation circuit to prevent high-voltage misconnection.
  • Supports power-based meter testing function, can directly read the power measurement value of the tested meter via 485 bus interface and calculate error.
  • Can display parameters such as power, voltage, current, and ripple.
  • Supports PC connection; dedicated calibration software has a user-friendly interface and simple, intuitive calibration settings. Users can manually add and set any calibration points, with data storage and printing capabilities.
  • Equipped with GPS clock reference, supporting daily timing error testing of energy meters.
  • Supports self-heating test, allowing self-heating influence experiments on the tested meter and providing error variation curves.
  • The device uses single-phase power supply for user convenience.
  • Each slot has automatic and manual switching capabilities, allowing users to select slots participating in calibration easily via software.
  • Supports barcode information scanning input for the tested meter; scanned information automatically associates with the tested meter (requires additional barcode scanner).
  • Equipped with 220VAC, 12VDC, 24VDC auxiliary power interfaces, adjustable ±20%, enabling power drop tests for energy meters.

 

Technical Specifications

 

DC Energy Meter Slots
DC Energy Meter Slots 18 slots
Eighteen-channel DC voltage output
Range 0~1500V (MAX: 1725V)
Ranges 100mV, 300mV, 1V, 3V, 10V, 30V, 100V, 200V, 300V, 1500V
Accuracy 0.012%RD + 0.008%RG
Output Power 200
10
U≤10 V: ≤ 10 mA;
One DC current output channel
Range 0~100A (MAX: 120A)
Ranges 1mA, 3mA, 10mA, 30mA, 100mA, 300mA, 1A, 3A, 10A, 30A, 100A;
Accuracy 0.012%RD + 0.008%RG
Load Capacity Port voltage: 8V
Output power: <800VA
Energy Accuracy
Energy Accuracy 0.05%
Voltage and Current Output Stability
Voltage Stability ±(0.01%*RD + 0.005%RG)/min
Current Stability ±(0.01%*RD + 0.005%RG)/min
Annual Drift Less than 100ppm/year
Voltage and Current Ripple Output Capability
Ripple Frequency 10~500Hz
Ripple Factor Voltage output allowed up to 10%
Current output allowed up to 20%
Ripple Phase Voltage and current ripple phase error less than 0.1 degrees
Ripple Accuracy Better than 5%
Auxiliary Power Supply
DC Output DC output: 12V, 24V
AC Output AC output: 220V
Adjustable Range ±20%
Output Power 80W
  Other Indicators
Operating Power Supply Power Supply Voltage: AC 220V±5% 50Hz
Power: Not greater than 2.2kW
Relative Humidity ≤70% No condensation, no corrosive gases
Temperature Operating Temperature: 15-35℃
Storage Temperature: -5℃ to 40℃
Total Weight Approximately 300kg
Dimensions Computer Desk: Length 723mm × Width 850mm × Height 770mm
Main Console: Length 1592mm × Width 814mm × Height 1333mm

Product Document

File Name

Update Time

Related Products

XL-9100T6 DC Meter Calibration Device


The XL-9100T6 DC energy meter calibration bench includes a standard source cabinet and an energy meter hanging table body. The cabinet houses the voltage source, small signal source, current source, main control unit, and industrial control computer display unit.

XL-9201 DC Shunt Calibrator


The XL-9201T3 DC shunt calibrator can test shunts of various specifications, featuring functions such as testing the basic error and DC resistance of shunts, plotting I-R and R-t curves, and automatically calculating the power factor. Its measurement accuracy is 0.05 level. This device is equipped with a high-precision, high-stability DC standard source and DC standard meter, along with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire testing process and generate a test report. The fixture can be adjusted to accommodate shunts of various specifications, providing high testing efficiency.

XL-9300T6 DC Energy Meter Calibration Device


The XL-9300T6 includes a voltage source, current source, small signal source, main control unit, and meter mounting platform. It can output 0 to 1000V DC (maximum 1150V) direct current voltage, 0 to 300A DC direct current output, and 0 to 5V DC small signal output. It is capable of calibrating both indirect and direct DC energy meters with an energy accuracy level of 0.05. The meter mounting platform is mainly used for securing meter positions, featuring 6 slots, allowing simultaneous calibration of 6 energy meters.

XL-9600T6 DC Meter Calibration Device


The XL-9600T6 DC meter calibration device includes a standard source cabinet and an electricity meter hanging table body. The cabinet houses a voltage source, a small signal source, a current source, a main control unit, and an industrial control computer display unit. It can output DC voltage from 0 to 1000V (MAX: 1150V), DC current from 0 to 600A, and small signals from 0 to 5V, enabling incremental output for electricity meter calibration. The voltage accuracy is 0.02 level, current accuracy is 0.02 level, and energy accuracy is 0.05 level. The hanging table body is mainly used for meter position fixing, including a meter rack and table body, with six meter positions, capable of simultaneously calibrating 6 electricity meters (0~100A) and 3 electricity meters (0~600A).

XL-9201T6 DC Shunt Calibrator


The XL-9201T6 DC shunt calibrator can test various specifications of shunts according to the JJG1069-2011 DC shunt calibration regulations. It is capable of testing the basic error and DC resistance of shunts, plotting I-R curves and R-t curves, among other functions. Its measurement accuracy level is 0.05. This device is equipped with a built-in high-precision, high-stability DC standard source and DC standard meter, along with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire testing process and generate a test report. The fixture can be adjusted to accommodate various specifications of shunts, providing high testing efficiency.

XL-9000T8 DC Meter Calibration Device


The XL-9000T8 DC energy meter calibration device includes a voltage source, a small signal source, a main control unit, and a meter mounting platform. It can output DC voltage from 0 to 1000V (MAX: 1150V) and a small signal of 0 to 5V, enabling the calibration of indirect DC energy meters. The voltage accuracy is 0.02 level, and the energy accuracy is 0.05 level. The meter mounting platform is mainly used for fixing meter positions, featuring 8 slots, allowing simultaneous calibration of 8 energy meters.

XL-9000T16 DC Meter Calibration Device


The XL-9000T16 DC energy meter calibration device includes a voltage source, a small signal source, a main control unit, and a meter mounting platform. It can output 0 to 1000V (MAX: 1150V) DC voltage and 0 to 5V small signals, enabling the calibration of indirect DC energy meters. The voltage accuracy is 0.02 level, and the energy accuracy is 0.05 level. The meter mounting platform is mainly used for fixing meter positions, featuring 16 slots, allowing simultaneous calibration of 16 energy meters.

Product Inquiry

Our staff will contact you within 24 hours (working days)