XL-9100T18 DC Meter Calibration Device
The XL-9100T18 DC energy meter calibration device is equipped with 18 independent voltage sources and 1 current source. It can output DC voltage from 0 to 1500V (MAX: 1725V) and DC current from 0 to 100A, enabling incremental output calibration for direct DC energy meters. The voltage accuracy is 0.02 level, current accuracy is 0.03 level, and energy accuracy is 0.05 level. It features a dedicated meter rack with 18 slots, allowing simultaneous calibration of 18 direct energy meters (0~100A).
Product Description
Product Introduction
XL-9100T18 DC Energy Meter Calibration Device Built-in 18 independent voltage sources and 1 current source, capable of outputting 0~1500V (MAX: 1725V) DC voltage and 0~100A DC current, enabling incremental output for the calibration of direct DC energy meters. Voltage accuracy is 0.02 level, current accuracy is 0.03 level, and energy accuracy is 0.05 level; designed with a dedicated meter rack with 18 slots, allowing simultaneous calibration of 18 direct energy meters (0~100A).
Main Features
- The enclosure is made of special aluminum alloy material, elegant and sturdy in structure, lightweight, and easy to transport.
- Equipped with a dedicated meter rack for convenient meter calibration.
- High-quality linear power amplifier with low ripple output and good stability.
- Multiple protection systems with audible and visual alarm functions for higher reliability.
- The operation software includes shortcut keys for common calibration points, allowing quick setting of frequently used test points during manual meter calibration, making manual calibration more convenient.
- Equipped with RS485 communication port, capable of communicating with the tested DC meter to read electrical parameters.
- Fully automatic testing of meter start, creep, basic error, daily timing error, constant test, etc., according to regulations.
- Supports ripple influence test; voltage and current can superimpose 10~500Hz ripple, with ripple factor and phase adjustable as needed.
- Dedicated pulse sampling interface with high-voltage isolation circuit to prevent high-voltage misconnection.
- Supports power-based meter testing function, can directly read the power measurement value of the tested meter via 485 bus interface and calculate error.
- Can display parameters such as power, voltage, current, and ripple.
- Supports PC connection; dedicated calibration software has a user-friendly interface and simple, intuitive calibration settings. Users can manually add and set any calibration points, with data storage and printing capabilities.
- Equipped with GPS clock reference, supporting daily timing error testing of energy meters.
- Supports self-heating test, allowing self-heating influence experiments on the tested meter and providing error variation curves.
- The device uses single-phase power supply for user convenience.
- Each slot has automatic and manual switching capabilities, allowing users to select slots participating in calibration easily via software.
- Supports barcode information scanning input for the tested meter; scanned information automatically associates with the tested meter (requires additional barcode scanner).
- Equipped with 220VAC, 12VDC, 24VDC auxiliary power interfaces, adjustable ±20%, enabling power drop tests for energy meters.
Technical Specifications
DC Energy Meter Slots | |
DC Energy Meter Slots | 18 slots |
Eighteen-channel DC voltage output | |
Range | 0~1500V (MAX: 1725V) |
Ranges | 100mV, 300mV, 1V, 3V, 10V, 30V, 100V, 200V, 300V, 1500V |
Accuracy | 0.012%RD + 0.008%RG |
Output Power | 200 |
10 | |
U≤10 V: ≤ 10 mA; | |
One DC current output channel | |
Range | 0~100A (MAX: 120A) |
Ranges | 1mA, 3mA, 10mA, 30mA, 100mA, 300mA, 1A, 3A, 10A, 30A, 100A; |
Accuracy | 0.012%RD + 0.008%RG |
Load Capacity | Port voltage: 8V |
Output power: <800VA | |
Energy Accuracy | |
Energy Accuracy | 0.05% |
Voltage and Current Output Stability | |
Voltage Stability | ±(0.01%*RD + 0.005%RG)/min |
Current Stability | ±(0.01%*RD + 0.005%RG)/min |
Annual Drift | Less than 100ppm/year |
Voltage and Current Ripple Output Capability | |
Ripple Frequency | 10~500Hz |
Ripple Factor | Voltage output allowed up to 10% |
Current output allowed up to 20% | |
Ripple Phase | Voltage and current ripple phase error less than 0.1 degrees |
Ripple Accuracy | Better than 5% |
Auxiliary Power Supply | |
DC Output | DC output: 12V, 24V |
AC Output | AC output: 220V |
Adjustable Range | ±20% |
Output Power | 80W |
Other Indicators | |
Operating Power Supply | Power Supply Voltage: AC 220V±5% 50Hz |
Power: Not greater than 2.2kW | |
Relative Humidity | ≤70% No condensation, no corrosive gases |
Temperature | Operating Temperature: 15-35℃ |
Storage Temperature: -5℃ to 40℃ | |
Total Weight | Approximately 300kg |
Dimensions | Computer Desk: Length 723mm × Width 850mm × Height 770mm |
Main Console: Length 1592mm × Width 814mm × Height 1333mm |