XL-9201 DC Shunt Calibrator

The XL-9201T3 DC shunt calibrator can test shunts of various specifications, featuring functions such as testing the basic error and DC resistance of shunts, plotting I-R and R-t curves, and automatically calculating the power factor. Its measurement accuracy is 0.05 level. This device is equipped with a high-precision, high-stability DC standard source and DC standard meter, along with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire testing process and generate a test report. The fixture can be adjusted to accommodate shunts of various specifications, providing high testing efficiency.

Product Description

Product Introduction

XL-9201 Shunt Calibrator Device Can test shunts of various specifications, with functions including testing basic error, error consistency, error variation, repeatability, thermal equilibrium time, DC resistance, plotting I-R and R-t curves, etc. Measurement accuracy is 0.05 level.
This device has a built-in high-precision, high-stability DC standard source and DC standard voltmeter, equipped with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire process and generate a test report. The fixture is adjustable to fit various shunt specifications, providing high testing efficiency.

 

Function Introduction

 

  • Basic error calibration of shunts, error consistency, error variation, repeatability test, thermal equilibrium time test, overload test;
  • Calculation of shunt DC resistance;
  • Measurement of I-R variation curve;
  • Power factor calculation;
  • Automatic report generation;
  • Historical calibration data retrieval;
  • Automatic calibration, supports one-key completion of shunt calibration after wiring.

 

Main Features

 

  • Efficient testing platform based on high-speed DSP, large-scale FPGA, high-speed high-precision DA converter, and high-performance power amplifier;
  • Output accuracy within 0.03%;
  • Current output 0~600A, stability within 0.01%;
  • Convenient calibration plan editing function, plans can be saved and recalled;
  • Semi-automatic testing, one-key completion of shunt testing after wiring;
  • Equipped with fixtures supporting rapid shunt calibration;
  • Connected to host computer via network port for high work efficiency.

 

Technical Specifications

 

Name Specification Range
Number of Phases Single-phase Current
DC Output Range 0~600A
Current Stability 0.01%/1 minute
DC Output Levels 1mA, 5mA, 20mA, 50mA, 200mA, 1A, 5A, 10A, 15A, 60A, 100A, 500A
Adjustment Resolution 0.001% RG
Output Stabilization Time Stabilization time to specified accuracy ≤1ms
Accuracy 0.02% RD + 0.01% RG
Annual Drift 60ppm
Resolution Less than 1uA
Load Capacity Range 1mA~20A: 3V;
Range 20A~200A: 2.5V
Range 600A: 2V
DC Source Ripple Factor Less than 1%
Measurement Meter Technical Specifications
Accuracy ±(% reading + % range)
Range 24 hours (23±1℃) 90 days (23±5℃) 1 year (23±5℃) Temperature Coefficient
1mV 0.01 + 0.002 0.01 + 0.003 0.015 + 0.003 0.0004 + 0.0001
10mV 0.01 + 0.0002 0.01 + 0.0003 0.015 + 0.0004 0.0004 + 0.0001
100mV 0.008 + 0.003 0.009 + 0.0005 0.008 + 0.0005 0.0004 + 0.00006
1V 0.008 + 0.0003 0.009 + 0.0005 0.01 + 0.0005 0.0004 + 0.00004
10V 0.008 + 0.0001 0.009 + 0.0005 0.01 + 0.0005 0.0001 + 0.00002
100V 0.008 + 0.0004 0.009 + 0.0006 0.01 + 0.0006 0.0004 + 0.00005
  Other Parameters
Operating Voltage AC 220V±5% 50Hz
Total Power Consumption No load: 120W, full load: 1.9KW
Warm-up Time <15 minutes
Operating Temperature -10℃~55℃
Relative Humidity ≤85% Non-corrosive gases
Dimensions 550mm*498.5mm*948mm (L*W*H)
Total Weight 100KG

 

Product Document

File Name

Update Time

Related Products

XL-9100T6 DC Meter Calibration Device


The XL-9100T6 DC energy meter calibration bench includes a standard source cabinet and an energy meter hanging table body. The cabinet houses the voltage source, small signal source, current source, main control unit, and industrial control computer display unit.

XL-9300T6 DC Energy Meter Calibration Device


The XL-9300T6 includes a voltage source, current source, small signal source, main control unit, and meter mounting platform. It can output 0 to 1000V DC (maximum 1150V) direct current voltage, 0 to 300A DC direct current output, and 0 to 5V DC small signal output. It is capable of calibrating both indirect and direct DC energy meters with an energy accuracy level of 0.05. The meter mounting platform is mainly used for securing meter positions, featuring 6 slots, allowing simultaneous calibration of 6 energy meters.

XL-9600T6 DC Meter Calibration Device


The XL-9600T6 DC meter calibration device includes a standard source cabinet and an electricity meter hanging table body. The cabinet houses a voltage source, a small signal source, a current source, a main control unit, and an industrial control computer display unit. It can output DC voltage from 0 to 1000V (MAX: 1150V), DC current from 0 to 600A, and small signals from 0 to 5V, enabling incremental output for electricity meter calibration. The voltage accuracy is 0.02 level, current accuracy is 0.02 level, and energy accuracy is 0.05 level. The hanging table body is mainly used for meter position fixing, including a meter rack and table body, with six meter positions, capable of simultaneously calibrating 6 electricity meters (0~100A) and 3 electricity meters (0~600A).

XL-9201T6 DC Shunt Calibrator


The XL-9201T6 DC shunt calibrator can test various specifications of shunts according to the JJG1069-2011 DC shunt calibration regulations. It is capable of testing the basic error and DC resistance of shunts, plotting I-R curves and R-t curves, among other functions. Its measurement accuracy level is 0.05. This device is equipped with a built-in high-precision, high-stability DC standard source and DC standard meter, along with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire testing process and generate a test report. The fixture can be adjusted to accommodate various specifications of shunts, providing high testing efficiency.

XL-9100T18 DC Meter Calibration Device


The XL-9100T18 DC energy meter calibration device is equipped with 18 independent voltage sources and 1 current source. It can output DC voltage from 0 to 1500V (MAX: 1725V) and DC current from 0 to 100A, enabling incremental output calibration for direct DC energy meters. The voltage accuracy is 0.02 level, current accuracy is 0.03 level, and energy accuracy is 0.05 level. It features a dedicated meter rack with 18 slots, allowing simultaneous calibration of 18 direct energy meters (0~100A).

XL-9000T8 DC Meter Calibration Device


The XL-9000T8 DC energy meter calibration device includes a voltage source, a small signal source, a main control unit, and a meter mounting platform. It can output DC voltage from 0 to 1000V (MAX: 1150V) and a small signal of 0 to 5V, enabling the calibration of indirect DC energy meters. The voltage accuracy is 0.02 level, and the energy accuracy is 0.05 level. The meter mounting platform is mainly used for fixing meter positions, featuring 8 slots, allowing simultaneous calibration of 8 energy meters.

XL-9000T16 DC Meter Calibration Device


The XL-9000T16 DC energy meter calibration device includes a voltage source, a small signal source, a main control unit, and a meter mounting platform. It can output 0 to 1000V (MAX: 1150V) DC voltage and 0 to 5V small signals, enabling the calibration of indirect DC energy meters. The voltage accuracy is 0.02 level, and the energy accuracy is 0.05 level. The meter mounting platform is mainly used for fixing meter positions, featuring 16 slots, allowing simultaneous calibration of 16 energy meters.

Product Inquiry

Our staff will contact you within 24 hours (working days)