XL-9201T6 DC Shunt Calibrator
The XL-9201T6 DC shunt calibrator can test various specifications of shunts according to the JJG1069-2011 DC shunt calibration regulations. It is capable of testing the basic error and DC resistance of shunts, plotting I-R curves and R-t curves, among other functions. Its measurement accuracy level is 0.05. This device is equipped with a built-in high-precision, high-stability DC standard source and DC standard meter, along with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire testing process and generate a test report. The fixture can be adjusted to accommodate various specifications of shunts, providing high testing efficiency.
Product Description
Product Introduction
The XL-9201T6 DC Shunt Calibrator can test various specifications of shunts according to the JJG1069-2011 DC Shunt Calibration Procedure. It is capable of testing the basic error and DC resistance of shunts, plotting I-R and R-t curves, among other functions. Its measurement accuracy level is 0.05.
This device is equipped with a built-in high-precision, high-stability DC standard source and DC standard meter, along with auxiliary fixtures for shunt calibration. After editing the test plan and placing the shunt, starting the test will automatically complete the entire process and generate a test report. The fixture is adjustable to accommodate various shunt specifications, providing high testing efficiency.
Function Introduction
- Basic error calibration of shunts, error consistency, error variation, repeatability testing, thermal equilibrium time testing, overload test;
- Calculation of shunt DC resistance;
- Calibration of error consistency, error variation, error repeatability, thermal equilibrium time, etc.;
- Measurement of I-R and R-t variation curves;
- Automatic report generation;
- Historical calibration data retrieval;
- Automatic calibration, supporting one-click completion of shunt calibration after wiring.
Main Features
- The XL-9201T6 DC Shunt Calibrator is an efficient testing platform based on high-speed DSP, large-scale FPGA, high-speed high-precision DA acquisition, and high-performance power amplifiers;
- The standard source output accuracy of the XL-9201T6 DC Shunt Calibrator reaches within one ten-thousandth;
- Current output range is 0~600A, with stability within 0.005%;
- Convenient calibration plan editing function, with plans that can be saved and recalled;
- Automatic testing, allowing one-click completion of shunt testing after wiring;
- Equipped with fixtures supporting rapid shunt calibration.
- Connected to the host computer via network port for high work efficiency.
Technical Specifications
DC Source Technical Specifications | ||||
Serial Number | Name | Specification Range | Remarks | |
1 | Number of Phases | Single-phase Current | ||
2 | DC Output Range | 0~600A (up to 700A for short periods) | ||
3 | Current Stability | 0.005%/1 minute | ||
4 | DC Output Ranges | 1mA, 5mA, 20mA, 50mA, 200mA, 1A, 5A, 10A, 20A, 60A, 100A, 500A | Supports automatic range switching | |
5 | Adjustment Resolution | 0.001% RG | ||
6 | Output Stabilization Time | Stabilization time to specified accuracy ≤1ms | ||
7 | Accuracy | (0.01A | ||
8 | Annual Drift | 60ppm | ||
9 | Resolution | Less than 1uA | ||
10 | Load Capacity | 1mA to 20A range: 3V | ||
20A to 200A range: 2.5V | ||||
600A range: 3V | ||||
11 | DC Source Ripple Factor | Less than 0.5% | ||
Note: To ensure output accuracy, the output current must be kept within the limit; otherwise, accuracy requirements may not be met. | ||||
Measurement Meter Technical Specifications | ||||
Accuracy Specification ±(% reading + % range) | ||||
Range | 24 hours (23±1℃) | 90 days (23±5℃) | 1 year (23±5℃) | Temperature Coefficient |
1mV | 0.01 + 0.002 | 0.01 + 0.003 | 0.015 + 0.003 | 0.0004 + 0.0001 |
10mV | 0.01 + 0.0002 | 0.01 + 0.0003 | 0.015 + 0.0004 | 0.0004 + 0.0001 |
100mV | 0.008 + 0.003 | 0.009 + 0.0005 | 0.008 + 0.0005 | 0.0004 + 0.00006 |
1V | 0.008 + 0.0003 | 0.009 + 0.0005 | 0.01 + 0.0005 | 0.0004 + 0.00004 |
10V | 0.008 + 0.0001 | 0.009 + 0.0005 | 0.01 + 0.0005 | 0.0001 + 0.00002 |
100V | 0.008 + 0.0004 | 0.009+0.0006 | 0.01+0.0006 | 0.0004+0.00005 |
Other parameters | ||||
Operating voltage | AC 220V±5% 50Hz | |||
Total power consumption | No load: 120W, full load: 3kW | |||
Preheating time | <15 minutes | |||
Operating temperature | -10℃ to 55℃ | |||
Relative humidity | ≤85% Non-corrosive gases | |||
Dimensions | 1900mm*900mm*1150mm (L*W*H) |
Product Document